Grating | Tuning range | Linewidth | |
LIOPSTAR-HQ | 1800 l/mm, 90 mm | 430 nm – 900 nm | < 0.07 cm-1 @ 570 nm |
LIOPSTAR-HQ | 2400 l/mm, 90 mm | 430 nm – 750 nm | < 0.05 cm-1 @ 570 nm |
Nd:YAG
532 nm, 10 kHz, 10ns, 150W Nd:YAG 532 nm, 10 kHz, 10ns, 90W Nd:YAG 532 nm, 10 kHz, 10ns, 50W |
> 40
W @ 564 nm > 6.0 W @ 282 nm *
Rh6G > 25 W @ 564 nm > 3.5 W @ 282 nm * > 12 W @ 564 nm > 1.7 W @ 282 nm |
Nd:YAG 532 nm, 10 kHz, 10ns, 150W Nd:YAG 532 nm, 10 kHz, 10ns, 90W Nd:YAG 532 nm, 10 kHz, 10ns, 50W |
>
35 W @ 640 nm > 5.5 W @ 282 nm
* DCM > 25 W @ 640 nm > 3.5 W @ 282 nm * > 10 W @ 640 nm > 1.5 W @ 282 nm |
* with Boost option | |
Nd:YAG 355 nm, 10 kHz, 10ns, 70W Nd:YAG 355 nm, 10 kHz, 10ns, 50W |
>10.5
W @ 442 nm > 1.5 W @ 221 nm
C120 > 7.5 W @ 442 nm > 1.1 W @ 221 nm C120 |
Nd:YAG 355 nm, 10 kHz, 10ns, 70W Nd:YAG 355 nm, 10 kHz, 10ns, 50W |
>10.5
W @ 488 nm > 1.5 W @ 244 nm
C102 > 7.5 W @ 488 nm > 1.1 W @ 244 nm C102 |
wavelength reproducibility | < 0.005 nm |
absolute accuracy | < 0.01 nm |
scan linearity | < 0.002 nm |
wavelength stability | < 0.001 nm/°C |
divergence | 0.5 mrad |
polarisation | > 98 % vertical |
ASE-background | < 0.5 % |
Dimensions :
LIOPSTAR-HQ | 1040 mm x 400 mm x 300 mm ± 10 mm, 80 kg |
Beam input height | 180 mm ± 10 mm |
Beam output height | 200 mm ± 10 mm |
Requirements :
Cooling for dye solvent | 800 Watt, resonator & amplifier system |
Laboratory | dust-free air (flow box) |
Voltage | 110...230V, single phase, 50 Hz/ 60 Hz |
Computer | single USB port |
Operating system | Windows XP/ Windows Vista/ Windows 7 |
Pump laser pulse power, kHz operation | up to 150 Watt |
Specification are subject to change without notice
Due to the amplifier dye cell set up in Brewster angle, the reflections and the parasitic lasing are minimized.
This leads to an overall higher performance of the laser output.
Pulsed dye laser Datasheet (pdf / English)